Product Description
Product details
| Meter Type | Conductivity |
| Measurement Range | 0.00 19.99 ms/cm |
| Accuracy | 2.5% FS |
| Item Number: | 30073974 |
Specification
| Meter Type | Conductivity |
| Measurement Range | 0.00 19.99 ms/cm |
| Accuracy | 2.5% FS |
| Measurement Resolution | 10 s/cm |
| Display | LCD |
| ATC | Not Applicable |
| Temperature compensation | Not applicable |
| Automatic Shutdown | Yes, After 6 minutes of non-use |
| Battery Life | 200 Hours with Disposable Batteries |
| Calibration | NA |
| Conductivity accuracy () | 2.5 % |
| Conductivity measuring range | 0.00 19.99 mS/cm |
| Dimensions | 1.5 in x 7.28 in x 1.77 in (38 mm x 185 mm x 45 mm) (LxHxW) |
| Net Weight | 3.88 oz (110 g) |
| Number of Buttons | 1 |
| Power | 4 AG13 Batteries (Included) |
| Protection Rating | IP67 |
| Temperature Display | Not Applicable |
| Working Environment | 50F 104F, 85%RH, non-condensing (10C 40C, 85% RH, non-condensing) |
| Wrist Strap | Yes |
Precision Measurement for LaboratoriesEngineered for laboratory use, the ST10C-C Starter Pen Meter provides consistently accurate conductivity readings. Its wide measurement range and high resolution make it suitable for various dissolved solids and conductivity applications. The robust design ensures dependable performance even in demanding work environments.
User-Friendly and Portable DesignThe ST10C-C features a compact form factor, lightweight build, and a single intuitive button for operation. The durable construction, along with an IP67 protection rating, means it can withstand exposure to water and dust, perfect for busy professionals. The included wrist strap enhances portability and security during use.
FAQ's of ST10C-C Starter Pen Meters:
Q: How do I operate the ST10C-C Starter Pen Meter?
A: To operate the ST10C-C, simply insert the included AG13 batteries, power on the device using its single button, and submerge the probe in the sample solution. The LCD will display the conductivity measurement instantly.
Q: What conditions are ideal for using this conductivity meter?
A: The ideal working environment is between 10C and 40C with a relative humidity of up to 85%. This ensures optimal measurement accuracy and device longevity.
Q: When should I replace the batteries in the meter?
A: Replace the AG13 batteries when the display dims or the meter fails to power on. With a typical battery life of 200 hours, frequent users can expect extended periods between changes.
Q: Where can the ST10C-C be used?
A: It is designed for laboratory environments but its waterproof and dustproof rating (IP67) also allows use in field settings, water quality studies, and various industrial applications.
Q: What process should I follow to maintain the meter?
A: After each use, rinse the probe with distilled water and gently dry it. Store the meter in a clean, dry place, and avoid exposure to extreme temperatures or humidity outside the recommended range.
Q: What are the key benefits of using this pen meter?
A: Users benefit from its portability, high accuracy, and ease of operation. The rugged build and waterproof rating make it reliable for consistent results in diverse conditions.